Product List Home Main Product Plant Extract Natural Ingredients Solution Contact Us Exhibition
About staherb
Home
News
Support & Downloads

News 

News

Asiaticoside and Madecasoside by HPLC

Publish Date 2019-12-02
Determination of Asiaticoside and Madecasoside by HPLC-ELSD

1) Reagents
a. Nanopure water b. Acetonitrile (HPLC)

2) Apparatus
HPLC (Shimadzu) Analytical balance (1/10000) Ultrasonic cleaner
0.45μm PTFE filter 50ml volumetric flask Screw cap vials
ELSD detector(Alltech)

3) HPLC condition
Column: Kromasil C18 5 m 4.6200mm
Mobile Phase: A:Water B: Acetonitrile
A:B=67:33
Flow Rate: 0.7ml/min
Injection: 10μl
Drift tube temperature :100.2℃
Gas flow:2.7L/min
Impactor: Off
Gain:2

4) Procedure
Samples Preparation
Accurately weigh 30mg extraction into 50ml volumetric flask. Add Water to volume and sonicate to mix, filter through a 0.45μm PTFE filter into a screw cap vial and proceed with the analysis.

Standard preparation /calibration
Prepare a standard solution (approximately 0.2mg/ml in Water using a standard of Asiaticoside and Madecasoside ,Sonicate to mix, filter through a 0.45μm PTFE filter into a screw cap vial and proceed with the analysis.

5) Calculation
Determine the peak area’s logarithm of Asiaticoside and Madecasoside the on the sample chromatogram and compare to the corresponding peaks in the chromatogram for the standard solution.



Product List  |  Home  |  Main Product  |  Plant Extract  |  Natural Ingredients  |  Solution  |  Contact Us  |  Exhibition  |  Sitemap  |  Mobile Version
  English     简体版     繁體版
HomeContact UsSitemap